Phosphate Testing Procedure
- Fill
three test tubes to mark with sample water. (Remember to read at the bottom of the meniscus.)
- Treat
one of the test tubes according to test procedures. (The other two will be used as
blanks)
- Use
1.0 mL pipet to add 1.0 mL of Phosphate Acid Reagent.
- Cap
and mix. (Note: The cap will
pop off if you do not hold it on while mixing. Shake gently to
mix and make sure you are wearing gloves.)
- Use
0.1 g spoon to add one level measure of Phosphate Reducing Reagent.
- Cap
and mix until dissolved.
- Remove
Stopper from Test tube.
- Insert
test tube with treated sample into slot in Axial Reader directly behind
distilled water ampoule.
- Insert
two test tubes with untreated sample into slots in Axial Reader on either
side of treated sample.
- Slide
Octect Comparator down until top is even with top of Axial Reader. (Hold comparator so natural light
shines down through test tubes.)
Compare color in center test tube to colors in top left corner of
Octet Comparator.
- If
the color of the test sample is less than the color of the lowest value,
the result is recorded as ̉less thanÓ the lowest value.
- If
the color of the test sample matches one of the color standards in the
upper left-hand quadrant, the result is taken as the value of that color
standard.
- If
the color of the test sample falls between these two values, the result
is the average of these two values.
- If
the color of the test sample is darker than the color of the second color
standard, move the comparator to a position where the bottom of the Axial
Reader and the bottom of the comparator are even.
- If a
color match is not reached with the standards on the left-hand side of
the comparator, the test sample and blank tubes are transferred to the
right-hand side of the Axial Reading device. The ampoule of distilled water is transferred to the
hole on the right-hand side of the comparator.